Electrosurgical Generator

Performance features:

Cut Mode

Pure Cut

A high-frequency alternating current yields smooth, rapid cuts that evoke little to know hemostasis.

Blend 1

Preset blended cut modes give the surgeon varying degrees of hemostasis in cut mode.

Blend 2

The mode is used for vaporization of prostatic tissue or any fat tissue. The higher output wattage gives fast cutting effect of the tissue.

Blend 3

It is used for major coagulation in major surgery, with some cutting functions.

Endo Cut

The mode is used to automatically controlled cut system to reduce the complication rate of endoscopic sphincterotomy (EST) and serum hyperamylasemia after EST compared to the conventional blended cut mode.

Contact Now E-Mail Telephone
Product Details

CM-350A meets the surgeon's needs in the wide range of general procedures and in special ties including Gynecology, Urology, Neurology, Thorasic, plastic and Reconstructive surgery. 

It has spray coagulation which rivals the best spark gap generators and with minimum cutting effect. The blend gives a good Hemostasis effect than cutting and pure cut is exceptionally smooth and starts promptly even in irrigated producers.

The isolated bipolar output, which has a non-sparking characteristic is ideal for microsurgery, neuro surgery laparoscopy and other applications.


CM-350A Electrosurgical Generator  CM-350A Electrosurgical Generator CM-350A Electrosurgical Generator 


Coag Mode

Spray 

Spray output is made for direct coagulation at the tissue with spark. above spark does not give any cutting effect as its CREST FACTOR is high.

Forcce

This mode is greater degree of homeostasis and covers all demands of standard coagulation. This produce the fine attune waveform that creates superficial resistance in the tissue resulting in superior coagulation, where the tissue destroyed by hyper evaporation. This give surgeons the ability to remove surface structures without damaging deeper or peripheral tissues and to cauterize large surface areas quickly when required.

Desicate

Coagulation in which the active electrode is in direct contact with the tissue is referred to as desiccation and is the type of coagulation used in most surgeries it ensures pinpoint desiccation with less destruction of peripheral tissue.


CM-350A Electrosurgical Generator

BipolarI Mode

Micro

High voltage force coagulation for special purpose when speed is required.

Macro

For highly precise control over coagulation process and privations of carbonization.

Bi-cut

This bipolar cut technology nearly eliminate excessive current even during precision cutting in minute tissue structure

Auto

Coagulation start by pressing the footswitch and stops output and audio signal automatically when the tissue is coagulated.


CM-350A Electrosurgical Generator


Rated output power:

Cut Mode

Pure cut: 1W~350W (Max. 400W, load 500 Ω); 

Blend cut 1: 1W~250W (load 500 Ω); 

Blend cut 2: 1W~200W (load 500 Ω); 

Blend cut 3: 1W~150W (load 500 Ω); 

Endo Cut: 99% 

Coag Mode

Spray coagulation: 1W~120W (load 500 Ω); 

Force coagulation: 1W~120W (load 500 Ω); 

Desicate coagulation: 1W~150W (load 500 Ω);

Bipolar 

Micro: 1W~80W (load 100 Ω);

Macro: 1W~80W (load 100 Ω);

Bi-cut: 1W~120W (load 100 Ω);

Auto: 1W~120W (load 100 Ω);


CM-350A Electrosurgical Generator


Power consumption of the whole machine: ≤ 1100VA. (cut function 350W);


CM-350A Electrosurgical Generator